2 Measuring and configuration
|
2.3 Electrical system
▶ Screw the compressed air cylinder back on.
▶ Perform the energy adjustment and “safety check“.
2.3.3 Adjust the KEY 3 Diagnodent electronics when the distance
standard is mounted
When one of the following components are exchanged, they must all be electrically
adjusted to each other: the KEY 3 Diagnodent electronics (Mat. no. 1.000.4271),
VEM distributor energy measuring electronics (Mat. no. 1.000.7724), adapter pho‐
tosensor (Mat. no. 1.000.4249) and pilot laser (Mat. no. 1.000.5631).
This is done when the distance standard is mounted (Mat. no. 1.001.2914).
161/176