![KaVo KEY Laser III 1243 Скачать руководство пользователя страница 44](http://html.mh-extra.com/html/kavo/key-laser-iii-1243/key-laser-iii-1243_technicians-instructions_658195044.webp)
Technician's instructions KEY Laser III 1243
2 Measuring and configuration
|
2.1 Control unit service program
Clear counters/op. period
This command is used to clear the counters and operating period.
Note
The pulse counter serves to document operating data of important KEY laser com‐
ponents. Only delete the pulse counter after express instructions from KaVo.
Confirm the command Clear counters/op. period with OK, and the following safety
query appears:
If you confirm the safety query with YES, the pulse counter and operating hours are
deleted SINCE OPERATING PERIOD. The date SINCE OPERATING DATE is re‐
placed with the current date.
In the Operating control menu, the confirmation appears: COUNTERS AND OP:
PERIOD RESET OK 1.
42/176