Technician's instructions KEY Laser III 1243
2 Measuring and configuration
|
2.2 Optical system
▶ First lightly tighten the two outer screws and move the holder to the right, and
screw in the middle screw.
▶ Tighten all three screws.
▶ To make subsequent optical adjustment easier, mount the HR and PR mirror
holders parallel to the pump chamber.
▶ Then perform a complete adjustment according to these instructions.
2.2.4 Deflection mirror
Check the surfaces of the deflection mirror
▶ Switch off the KEY Laser.
▶ Remove the hood and cover of the optical system.
94/176