Technician's instructions KEY Laser III 1243
3 Maintenance work on KEY Laser 1243
|
3.7 Laser handpiece E 2062
3.7 Laser handpiece E 2062
Component
Jobs to be done
Exchange interval
Parts to replace as needed
Remove fibre plug and un‐
screw grip sleeve.
Exchange O-ring in the
grip sleeve.
O-ring
Mat. no. 0.200.6073 ➊
Replace O-ring on instru‐
ment inner part.
O-ring
Mat. no. 0.200.6220 ➊
Exchange O-ring on plug. O-ring
Mat. no. 0.200.6120 ➊
Visually check all lenses
and clean if necessary ac‐
cording to specifications
Lens adjusted
Mat. no. 1.001.0848
Check all available fibre
plugs for where he and the
damage and exchange if
necessary
Fibre plug 50/10
Mat. no. 0.574.0531
Fibre plug 30/28
Mat. no. 0.574.0501
Fibre plug 40/28
Mat. no. 0.574.0511
Fibre plug 50/28
Mat. no. 0.574.0521
Check silicon hose on all
fibre plugs for contamina‐
tion and wear an ex‐
change if necessary.
Hose, 16 mm long (1X
Mat. no. 0.574.2142 = 20
unit package)
Check cooling air and
transmission.
For details, see instructi‐
ons for use of the handpie‐
ces.