Mirus for H2 Classic GrainGage
58
H2 Classic GrainGage Calibration
29. Tap the check icon to save the new moisture curve.
If desired, first tap the gear icon on the right for more information about this moisture curve.
Note: Only adjust these settings if a HarvestMaster Field Service Engineer tells you to.
4.4.5 Test and Tune a Moisture Curve
Note: This section is only applicable to the EM moisture curve.
HarvestMaster recommends testing the new moisture curve against other grain samples to confirm the
moisture and test weight values.
1. Test the grain sample in a benchtop sensor by cycling sub-samples from the larger sample 3–5 times
and averaging the moisture and test weight readings. The averages are the Known Moisture
and
Test
Weight.
2. Within the hour, take the sample out to the GrainGage. If you are calibrating outside, make sure the
sample stays in the shade.
3. Open Mirus.
4. Go to
Setup
>
H2 GrainGage
>
Moisture Curves
.
5. Select the moisture curve you want to test.
6. Tap the edit icon .
Gear icon
Содержание HarvestMaster mirus H2
Страница 1: ......
Страница 7: ...CHAP TER ONE Mirus Ins tallation...
Страница 11: ...Mirus and GrainGage S etup CHAP TER T WO...
Страница 32: ...H2 Clas sic GrainGage Diagnos tic s and Aler t s CHAP TER THREE...
Страница 42: ...GrainGage Calibration CHAP TER FOUR...
Страница 71: ...Field Maps CHAP TER FIVE...
Страница 98: ...Har ves t Mode CHAP TER SIX...
Страница 111: ...Mirus for H2 Classic GrainGage 111 Harvest Mode 6 3 7 List View In List View you can see the data in the form of a chart...
Страница 120: ...Standard Grain Information APPENDIX A...
Страница 123: ...Troubleshooting APPENDIX B...
Страница 129: ...Limited Warrant y APPENDIX C...