Appendix B: Troubleshooting Mirus
8.3.2 Not Meeting Minimum Weight Threshold
The minimum weight threshold is the minimum amount of grain required for accurate moisture and test
weight data. By default, if the minimum weight threshold is not met, moisture and test weight data will be
recorded as zero.
To ensure the plots are yielding enough grain to measure accurate moisture and test weight,
1. Go to
Setup
>
Info
>
Chamber Info
.
•
If the minimum weight threshold value is outside
2–6 lb., recalibrate the test weight for the chamber
calibration.
•
If you are not harvesting enough grain to meet
the minimum weight threshold, install low yield
inserts and recalibrate the chamber. For more
information about low yield inserts, see
A: Standard Grain Information on page 121
•
If installing low yield inserts doesn’t work, there
is an option to override the minimum weight
threshold.
CAUTION: If the minimum weight threshold is
overridden, the moisture and test weight is not
guaranteed to be accurate.
To override the minimum weight threshold,
1. Go to
Setup
>
System
.
2. Change the Override Threshold option to
Yes
.
3. Tap
Save
.
!
Save
Содержание HarvestMaster mirus H2
Страница 1: ......
Страница 7: ...CHAP TER ONE Mirus Ins tallation...
Страница 11: ...Mirus and GrainGage S etup CHAP TER T WO...
Страница 32: ...H2 Clas sic GrainGage Diagnos tic s and Aler t s CHAP TER THREE...
Страница 42: ...GrainGage Calibration CHAP TER FOUR...
Страница 71: ...Field Maps CHAP TER FIVE...
Страница 98: ...Har ves t Mode CHAP TER SIX...
Страница 111: ...Mirus for H2 Classic GrainGage 111 Harvest Mode 6 3 7 List View In List View you can see the data in the form of a chart...
Страница 120: ...Standard Grain Information APPENDIX A...
Страница 123: ...Troubleshooting APPENDIX B...
Страница 129: ...Limited Warrant y APPENDIX C...