6.4.6 Accuracy of measurement
The measuring instrument has many measuring parameters, and the basic measuring parameters
are R,L,C,D
,
The rest of the parameters can be measured everywhere from the above parameters,
so the accuracy measurement mainly on R, L, C, D measurement.
6.4.7 Capacitance C, loss D accuracy
Function
Cp-D
Test frequency
100Hz 1kHz 10kHz 100kHz Separate test
Level
1V
Range
AUTO
Bias
0V
Speed
Slow
Short circuit and open circuit should be cleared before testing. Access standard capacitor
100pF,1000pF,10nF,0.1uF,1uF
,
Changing the frequency
,
The error capacitance C between
instrument reading and standard value shall be within the allowable error range stipulated in this
chapter for C accuracy, and the loss D shall be within the allowable error range stipulated in this
chapter for D accuracy.
6.4.8 Inductance L accuracy
Test conditions:
Function
Ls-Q
Test frequency 100Hz 1kHz Separate test
Level
1V
Range
AUTO
Bias
0V
Speed
Slow
Short circuit and open circuit should be cleared before testing. Access standard inductor
100μH,1mH,10mH,100mH
,
Changing the frequency,The error between the instrument reading
and the standard value shall be within the allowable error range stipulated in this chapter for L
accuracy.
6.4.9 Impedance Z accuracy
Test conditions:
Function
Z-θ
Test frequency
100Hz
,
1kHz
,
10kHz
,
100kHz Separate test
Level
1V