0mA—±50mA
,
Minimum resolution:5μA
30ΩInternal resistance:
0V— ±3V
Minimum resolution:0.5mV, Accuracy:1%x Set
5mV
0mA—±100mA
,
Minimum resolution:5μA
Accuracy of measurement
Measurement accuracy includes measurement stability, temperature coefficient, linearity,
repeatability and calibration interpolation error.
Examination of the accuracy of the instrument must be carried out under the following conditions:
a
.
Warm-up time: more than 30 minutes.
b
.
Warm-up time: more than 30 minutes.
c
.
Test cable length: 0m, 1M
d
.
Open circuit correctly and short circuit "0".after Preheating
e
.
The DC bias is in the "OFF" position, and the instrument range is at "AUTO" to select the
correct measurement range..
6.1.17 │Z│,│Y│,L,C,R,X,G,B accuracy
Ae is the accuracy of │Z│
,
│Y│
,
L
,
C
,
R
,
X
,
G
,
B expressed in the following way:
Ae = ±[A+(Ka+Kb+Kc)×100+ Kd +Kf]×Ke [%]
A:Basic measurement accuracy (see figure A)
Ka:Impedance scaling factor (see table A)
Kb:Impedance scaling factor (see table A)
Kc:Calibration interpolation factor (see table B)
Kd:Cable length factor (see table D)
Ke:Temperature factor (see table E)
Kf:Correction factor for scanning fixture(Not added:Kf = 0,added :Kf = 0.2)
L
,
C
,
X
,
B accuracy used condition: Dx (D measurement value) is≤0.1
R
,
G accuracy used condition:Qx(Q measurement value)≤0.1
When Dx≥0.1,four
L
,
C
,
X
,
B
the accuracy factor Ae should be multiplied by Root
Dx
2
When Qx≥0.1
,
For L, C, X, B, the accuracy factor Ae should be multiplied by Root number