35/55
soft key area.
l)
Press “OFF”
,
Turn off frequency 1 point frequency clearing function.
m)
Move the cursor to the frequency 2 area.“ON”,“OFF”,”Open circuit single frequency
clearing”
,
"Short circuit single frequency clearing "and" load correction"will be displayed in the
soft key area.
n)
Press “OFF”
,
Turn off frequency 2 point frequency clear function.
5.1.2 Point frequency zero(for better use of single frequency test):
Suppose the user now uses the test frequency of 5kHz.
a)
Press the menu key[SETUP]
,
Then press the soft key "user correction", the instrument will be
displayed as < user correction > page.
b)
Move the cursor to the open circuit area,“ON”,“OFF” and “open circuit full frequency
clearing”will be displayed in the soft key area.
c)
Press"ON" , turn on the open circuit correction function of the instrument.
d)
Moving cursor to short circuited area,“ON,“OFF”and“Short circuit full frequency clearing”
will be displayed in the soft key area.
e)
Press"ON",turn on the short circuit correction function of the instrument.
f)
Moving cursor to load area.“ON” , “OFF” will be displayed in the soft key area.
g)
Press "OFF" ,turn off the load correction function of the instrument.
h)
Moving cursor to frequency 1 area.“ON”, “OFF”,”Open circuit single frequency clearing”
,
"Short circuit single frequency clearing" and "load correction" will be displayed in the soft key
area
i)
Press “ON”
,
Turn on frequency 1 point frequency clearing function.
j)
Press[5]
,
5 will display the cursor area in the screen, and the software area will display the
available units(Hz,kHz,and MHz).press “kHz”,The frequency 1 area will be changed to
5.0000kHz (the same as the test frequency)
k)
Keep the test fixture open circuit
,
press “open-circuit single-frequency zero-clearing” to
perform open-circuit correction,
l)
Insert short circuit chip (JK26010) into the test fixture.
m)
Short circuit correction is performed according to soft key short circuit single frequency
clearing.
5.2 The correct connection of the tested components
The instrument has four pairs of test terminals: HCUR (current sampling high-end Hc), LCUR
(current sampling low-end Lc), HPOT (voltage sampling high-end Hp), LPOT (voltage sampling