45/55
The measured value is calculated for the absolute value deviation ΔABS of a programmable
nominal value and the percentage deviation of delta Δ%.
6.1.2 Equivalent method
Series and parallel connection
6.1.3 range
Automatic and manual (keep, increase or decrease)
6.1.4 trigger
Internal, external, manual
Internal: continuously measuring the measured parts and displaying the results.
Manual: press the panel "TRIGGER" key instrument for a measurement and output the results
show that normally in a waiting state.
External: The instrument HANDLER interface receives the "start" signal from the outside, makes
a measurement and outputs the measurement results, and then enters the waiting state again.
6.1.5 Trigger delay
Delay time: measure the time from trigger to start measurement.0 - 60 seconds, 1ms step
programmable
6.1.6 Test end connection mode
Four terminal measurement method is adopted.
Hcur: Current sampling high end
Lcur:Current sampling low end
Hpot:Voltage sampling high end
Lpot:Voltage sampling low end
6.1.7 Measuring speed (frequency >=10kHz )
Fast speed: about 30 times / sec (13ms/ times).
Medium speed: about 11 times / sec (90ms/ times).
Slow speed : about 2.7 times / sec (370ms/ times).
Medium speed and fast speed will decrease when the frequency is less than 10kHz.
6.1.8 Average
1 to 255 programmable.
6.1.9 Display digits
6 digits, maximum display number 999999
Test signal
6.1.10 Test signal frequency
The test signal is sine wave.Frequency accuracy:0.01
%
Test frequency range:
50Hz
~
100kHz(JK2817N)
20Hz
~
100kHz(JK2831)
20Hz
~
200kHz(JK2832)
Minimum resolution:0.01Hz