6.1.11 Signal mode
Normal: Set the test voltage on the measurement display page, and the voltage at the measurement
end may be smaller than the set voltage according to the impedance of the measured part.
Constant level: automatic adjustment of internal level makes the voltage on the tested part
consistent with the set voltage.
6.1.12 Test signal level
Mode
Range
Accuracy
Step
Volta
ge
Normal constant
pressure
5mVRMS—2VRMS
10mVRMS—5VRMS
±(10%×Set value+2mV)
±(6%×Set value+2mV)
100μV
Curr
ent
Normal constant
current
5OμARMS
—100mARMS
100μARMS
—50mARMS
±(10%×Set
value+10μARMS)
±(6%×Set
value+10μARMS)
1μA
6.1.13 Output impedance
30Ω,100Ω±2% Optional
6.1.14 Test signal level monitor
Mode
Range
Accuracy
Voltage
5mVRMS—2VRMS
0.01mVRMS—5mVRMS
±(3%×0.5mV)
±(12%×0.1mV)
Current
50μARMS—20mARMS
0.001μARMS—50μARMS
±(3%×5μA)
±(12%×1μA)
6.1.15 Measurement shows maximum range.
Parameter
Measurement display range
L,Lk
0.0001μH
~
99.999kH
C
0.0001pF
~
9.9999F
Z,R,X,DCR
0.0001Ω
~
99.999MΩ
Y,B,G
0.0001μs
~
99.999S
θ
Deg
-179.999°
~
179.999°
Rad
-3.14159
~
3.14159
D
0.0001 — 9.9999
Q
0.0001 — 9999.9
6.1.16 DC bias voltage source
100Ω Internal resistance: 0V— ±5V
Minimum resolution:0.5mV, Accuracy:1%x Set
5mV