When the reference element is connected to the tested end, press the soft key to
measure.JK2817N/JK2811B tests the reference components, and the test results are automatically
entered as reference A values.
2)Use soft keys to measure or use numeric keys to enter reference values of main parameters.
3)The reference value of the mobile cursor to enter the sub parameters in the reference B domain.
The soft key area of the screen displays the following soft keys.
●measurement
When the reference element is connected to the tested end, press the soft key to
measure.JK2817N/JK2811B tests the reference components, and the test results are automatically
entered as reference B values.
4)Use soft keys to measure or use numeric keys to input reference values of secondary
parameters.This step can be skipped if the reference values of the primary and secondary
parameters have been set in step 2 using the measurement soft key.
5)Move the cursor to the deviation A field. The soft key area of the screen displays the following
soft keys
●ABS
●%
●OFF
6)Use the above soft key to set the deviation of main parameters.
7)Move the cursor to the deviation B field. The soft key area of the screen displays the following
soft keys.
●ABS
●%
●OFF
8)Use the above soft key to set the deviation of auxiliary parameters.
3.4 < user correction > page
Press menu key [SETP], then press soft key user correction, enter user correction > page.
Figure:
The Open, Short Circuit, and Load Correction functions of the <User Correction> page can be
used to eliminate distributed capacitance, parasitic impedance, and other measurement