42/55
Main parameter(FUN1)
Cp
Auxiliary parameters(FUN2)
D
Frequency (FRQ)
100kHz
Level (LEV)
1V
Speed (SPEED)
SLOW(SLOW)
Auxiliary BIN switch(AUX)
ON
Main parameter tolerance mode(MODE)
%TOL(Percentage tolerance method)
Nominal value(NOMINAL)
270pF
BIN1 LOW (BIN1 LOW)
-4.6%
(BIN1 HIGH (BIN1 HIGH)
4.8%
BIN2 LOW (BIN2 LOW)
-9%
BIN2 HIGH(BIN2 HIGH)
10%
Auxiliary parameters LOW (2nd LOW)
0.0000
auxiliary parameters LOW(2nd LOW)
0.0015
TRIGGER MODE(TRIG)
EXT(external)
ALARM MODE(CMP ALARM)
Unqualified beep(Heigh long)
Explain 1:Because it is a small
capacitance
, its 100kHz impedance is greater than 1kΩ so we
choose the parallel equivalent mode
Explain 2:Capacity of any BIN qualified But when the loss is not qualified,Another BIN
processing.
,
So open the attachment BIN and classify it as an AUX BIN. If AUX is turned
off,When the loss is not qualified, the whole is judged to be unqualified.
Explain 3::Since the given upper and lower bounds are based on the percentage deviation of the
nominal 270pF value, the% TOL percentage mode is chosen as the main parameter.
Specific settings:
1) In component measurement display page (Meas), select Cp-D, set frequency, level, speed and
so on.
2) Press the SETUP menu key to enter the Meas Setup page and change the trigger mode to EXT
(external trigger).
3) Press SETUP-LIMIT to enter the limit list settings page(Limit Table), set nominal values, main
parameter tolerance mode, upper/lower limit parameters, comparator switch, auxiliary BIN
switch:
4) Press the SYSTEM menu key to enter the system settings page.(System Config)
,
Find the
unqualified beep item, set it to high long.
5) Return to component measurement display page after setup is completed.(Meas)
5.6.2Load calibration operation example
1 Operation steps:
Suppose that the test conditions that users are using now are as follows: