39/55
Press menu key [DISP],
JK2817N is displayed to < component measurement display > page.The instrument will be tested
continuously and the test results will be displayed in the center of the page.
7
.
If the test results are obviously wrong, please:
a)check whether the inductance is reliably connected with the test fixture.
b)check whether the test fixture is reliably connected with the test end of the instrument.
c) re open reliable open circuit / short circuit correction.
Note: When the user is using sweep open/short circuit correction, the point frequency correction
function must be selected as OFF, referring to the "zero-clearing correction operation" in this
chapter.
5.5 An example of fast operation of capacitance test with multi frequency list scanning using
JK2817N
The test conditions are as follows:
Function: Cp-D
Level : 1Vrms
Other parameters are as follows:
Frequency
Comparison parameters
Low
High
1kHz
Cp(capacity)
325.0nF
333.0nF
10kHz
D(loss)
0.0001
0.0003
100kHz
D(loss)
0.0060
0.0100
BEEP : HIGH LONG(High Long voice)
Alarm mode: OUT (ultra time)
The steps are as follows:
1. starting up, refer to the "starting up" section of the chapter "panel instruction and entry
operation".
2. basic parameter setting.
a)
Key [DISP]
,
JK2817N is displayed to < component measurement display > page.
b)
The current function area is Cp-D, and the level area is 1 V.
c)
Key [SETUP]
,
Make the instrument display to the <Measurement Settings> page, where the
“measurement settings”,” user correction”, “limit settings”, “list settings” and “file management”
will be displayed in the soft key area.
d)
Key ” list settings”
,
JK2817N is displayed to < list scan settings > page.
e)
Move cursor to scan parameter area. The current area is displayed as frequency [Hz].
f)
Press the knob key to move the cursor to the parameter area of scan point 1. The current area
is displayed as -.