Loss D known
10nF
0.1uF
1uF
2
AC Standard resistor
10Ω
0.02
%
100Ω
1kΩ
10kΩ
100kΩ
3
DC Standard resistor
0.1Ω
0.02
%
1Ω
10Ω
100Ω
1kΩ
10kΩ
100kΩ
4
Standard inductor
100μH
0.02%
1mH
10mH
100mH
5
Frequency meter
(0
~
1000)MHz
6
Digital multimeter
0.5
%
7
Insulation resistance meter
500V 10 level
8
Withstand voltage
leakage tester
0.25kW(0
~
500)V
6.4.3 Functional check
The function keys, monitors and terminals of the instrument should be able to function properly
and correctly.
6.4.4 Test signal level
A digital multimeter is placed in the AC voltage range, with one test bar connected to the HCUR
end of the meter and the other test bar to the ground end. The change level
is:10mV,20mV,100mV,200mV,1V,2V
,
The reading value should conform to the requirements of
the test signal level in this chapter.
6.4.5 Frequency
The ground terminal of the frequency meter is connected with the grounding end of the
instrument.The terminal of the frequency meter is connected with the HCUR end of the
capacitance tester.Changing the frequency:20Hz,100Hz,1kHz,10kHz,100kHz(JK2817N is
100kHz).The readings of the frequency meter should meet the requirements of the frequency of
the test signal in this chapter.