main parameter and the limit mode of the main parameter.
2)
Move the cursor to the lower limit setting area of BIN 1. If you choose the tolerance mode to
execute steps 3 to 6, if you choose the continuous mode to execute steps 7 to 11.
3)
Use the numeric key to enter the lower limit of BIN1 in the lower limit setting field of BIN1.
When data is entered, you can use the subscript soft key to determine the input data(p, n, µ, m, k,
M, *1)to determine the input data.When the soft key *1 is used, the limit value is F, H or Ω as the
default unit.After the limit value of the BIN1 is entered in the lower limit of BIN1,The lower limit
of BIN1 is automatically set to - (absolute limit),The upper limit of BIN1 is automatically set to +
(absolute limit).
4)
The cursor automatically jumps to the lower limit setting area of BIN2.Repeat step 4 until the
limit value of input BIN9.The cursor automatically jumps to the lower limit setting area of 2nd.
5)
After input the lower value limit of auxiliary parameters, the cursor will automatically jump to
the upper limit set area of 2nd.
6)
Input the upper limit of the auxiliary parameters.
7)
Use the numeric key to enter the lower limit of BIN1 in the lower limit setting field of BIN1.
When data is entered, you can use the subscript soft key to determine the input data(p, n, µ, m, k,
M, *1)to determine the input data.When the soft key *1 is used,the limit value is F, H or Ω as the
default unit.
8)
After input the lower value limit of BIN1,The cursor will automatically jump to the upper limit
setting area of BIN1 .Input the upper limit value of BIN1.
9)
The cursor will automatically jump to the upper limit setting area of BIN2.For continuous mode,
the lower limit of BIN2 is equal to the upper limit of BIN1.Input the upper limit value of BIN2.
10)
Repeat step 9 until the
upper limit value of input BIN9.The cursor automatically jumps to
the lower limit setting area of 2nd.Input the lower limit value of auxiliary parameters
11)
, the cursor will automatically jump to the upper limit set area of 2nd.Input the upper limit
value of auxiliary parameters