●Test function (function A)
●Test frequency (frequency)
●Test level (level)
●Test range (range)
●Test speed (speed)
3.2.1 Comparator function
JK2817N/JK2811B built-in comparison function can divide the tested components into up to 10
BINS.(BIN1 to BIN9 and BIN OUT).can set 9 pairs of main parameter BIN limits and a pair of
Auxiliary parameters BIN limits.If the main parameter of the part under test is within the BIN
limit, but the Auxiliary parameters is not within the BIN limit,The tested parts are sorted into the
auxiliary BIN.When JK2817N/JK2811B installs the HANDLER interface attachment, the
comparison test results can be output to the automatic test system for automatic sorting test. These
limit settings can only be set on the <Limit List Settings> page. The comparison field allows you
to set the comparison function ON or OFF.
Comparative function Operation steps:
1
)
Move the cursor to the comparison domain. The soft key area of the screen displays the
following soft keys.
●ON
●OFF
2)Use the soft key to set the comparison function to ON (open) or OFF (close).
3.2 < BIN count display > page
Press the menu key [DISP], then press the soft key [count[, enter the < BIN count display > page.
Display the count value of each file in the < BIN count display > page.
Figure:
The following measurement control parameters can be set in <BIN count display > page.
●Counting function ON/OFF (count)
This display page has two fields, they are: BIN count display, count. Each control function field
will be described in detail in the following paragraphs.
The test results / condition display area on this display page shows the following test condition
information.These surveillance domains look like settings, but they cannot be set on this display