43/55
Frequency: 100kHz
Cp standard value: 11nF
D standard value:0.0005
a) press key [SETUP]
,
Measurement settings, user correction, limit settings, list settings, file
management and tools are displayed in the soft key area.
b) press key”user correction”.The instrument will be displayed as < user correction > page.
c) Move the cursor to the open circuit.“ON”,”OFF”and “ON””Open circuit full frequency
clearing” will be displayed in the soft key area.
d) press key “ON”
,
Turn on the Open circuit calibration function of the instrument.
e) Move cursor to short circuit area.. “ON”,“OFF”and“Short circuit full frequency clearing”will
be displayed in the soft key area.
f)press key“ON”
,
Turn on the short circuit correction function of the instrument.
g) Move the cursor to the load area.“ON”
,
“OFF” will be displayed in the soft key area.
h)press key “ON”
,
Turn on the load correction function of the instrument.
i)Move the cursor to the functional area. The current area is displayed as Cp-D
,
At this
point,Cp-…
,
Cs-…
,
Lp-…
,
Ls-…
,
Z-…
,
More will be displayed in the soft key area on
the right side of the screen.
j)press key Cp-D select Cp-D parameter .k) Move the cursor to the frequency 1 area.“ON”
,
“OFF”
,
”Open circuit single frequency clearing”, “short circuit single frequency clearing” and
“load correction”will be displayed in the soft key area.
l)press ON key
,
Open the point frequency correction function of frequency 1.
m) press [1][0][0]
,
100 displayed on the screen, and the software area displays the available units
(Hz, kHz, and MHz).(Hz
,
kHz
,
and MHz).press kHz. frequency 1 area will be changed to
100.000kHz (the same as the test frequency).
n) Move the cursor to the reference A: area of frequency 1. Button [1][1] and 11 will be displayed
in the cursor area of the screen.And the software area will display the available units (p, n, mu, m,
k). The key n. The area will be changed to 11.0000nF.
o) Move the cursor to the reference B: area of frequency 1. Button[0][.][0][0][0][5]
,
0.0005 will be
displayed in the cursor area of the screen.And the software area will display the available units(p
,
n
,
μ
,
m
,
k).press [ENTER].The area will be changed to 0.00050.
Содержание JK2817N
Страница 1: ......