RS485 SERIAL PORT
72
HYDRO-VIEW USER GUIDE
D0022
REV 1.06
RS485 serial port
The RS485 serial port provides an optically-isolated twisted-pair interface between
multiple Hydro-View units and a host computer or master Hydro-View.
Installation of RS485 module
The RS485 interface is supplied on a plug-in module which must be inserted into the
Hydro-View unit unless this was installed as a factory option…
⇒
Remove the screws securing the lid of the Hydro-View.
⇒
Lift the lid off the unit to expose the circuit board.
⇒
Locate the RS485 module connector on the circuit board as shown in Figure 23 on
page 59.
⇒
Plug the module into this connector - only one orientation is possible.
⇒
Refit the lid and screws.
RS485 connections
The Hydro-View implements RS485 using a two wire interface. This is shown in Figure
32.
Shielded twisted-pair cable is recommended for EMC purposes. The shield should be
connected to ground at one end only to minimise capacitive coupling (the twisted pair
cable will minimise magnetic coupling).
RS485 termination
Note that the twisted pair cable should be terminated at each end in a resistor equivalent
to the characteristic impedance of the cable used - typically 100R/0.5W These are
indicated by Rt in Figure 32.
The RS485 module (shown in Figure 33) contains a termination resistor which may be
placed in circuit by inserting jumper J3 on the relevant module(s). This should be done for
those units at each end of the twisted pair.
•
Figure 32 - RS485 interface
Содержание Hydro-View HV02
Страница 20: ...20 HYDRO VIEW USER GUIDE D0022 REV 1 06 Notes ...
Страница 30: ...MATERIAL IN USE 30 HYDRO VIEW USER GUIDE D0022 REV 1 06 Notes ...
Страница 42: ...SUGGESTED WORKING VALUES 42 HYDRO VIEW USER GUIDE D0022 REV 1 06 Notes ...
Страница 48: ......
Страница 53: ...GROUPING CHANNELS HYDRO VIEW USER GUIDE D0022 REV 1 06 53 Figure 17 Four bin grouped system ...
Страница 54: ...EXAMPLE A FOUR BIN GROUPED SYSTEM USING RS485 54 HYDRO VIEW USER GUIDE D0022 REV 1 06 Notes ...
Страница 69: ...INSTALLATION HYDRO VIEW USER GUIDE D0022 REV 1 06 69 Notes ...
Страница 70: ......
Страница 80: ...ELECTROMAGNETIC COMPATIBILITY DIRECTIVE 89 336 EEC 80 HYDRO VIEW USER GUIDE D0022 REV 1 06 Notes ...
Страница 82: ...82 HYDRO VIEW USER GUIDE D0022 REV 1 06 Notes ...