ADVANCED USER OPTIONS AND CONFIGURATION
HYDRO-VIEW USER GUIDE
D0022
REV 1.06
33
Parameter
Default
Value
Description
+dV Coarse
1.00
Positive coarse filter input rate limit. Sets the maximum
positive voltage change between successive samples. Used
to filter out noise from the signal.
Range 0.01
Æ
10.00V
-dV Coarse
1.00
Negative coarse filter input rate limit. Sets the maximum
negative voltage change between successive samples. Used
to filter out noise from the signal.
Range 0.01
Æ
10.00V
+dV Fine
0.10
Positive fine filter input rate limit.
Range 0.01
Æ
10.00V
-dV Fine
0.10
Negative fine filter input rate limit.
Range 0.01
Æ
10.00V
Minimum
volts
0.5V
Minimum sensor voltage accepted before an alarm is issued.
Used to detect broken sensor cable.
Range 0.5V
Æ
Maximum volts.
Maximum
volts
12.0V
Maximum sensor voltage accepted before an alarm is
issued.
Range Minimum volts
Æ
12.0V
Trend step
1.0
Sets the update interval on the trend recorder display.
Range 0.1
Æ
10.0 seconds
Trend lower
0.00
Sets the minimum moisture value displayed on the trend
recorder display.
Range -99.99
Æ
99.99%
Trend upper 20.00
Sets the maximum moisture value displayed on the trend
recorder display.
Range -99.99
Æ
99.99%
Alarm Lo
0.00
Sets the minimum moisture value allowed before an alarm
occurs.
Range -99.99
Æ
99.99%
Alarm Hi
20.00
Sets the maximum moisture value allowed before an alarm
occurs.
Range -99.99
Æ
99.99%
Output set-up menu
The Hydro-View 0-10V (4-20mA) analogue output can be set to represent any convenient
moisture content range by choosing
Output set-up
from the
main menu
.
The parameters that may be set here are...
Parameter
Default
Value
Description
Minimum %
0.00
The moisture value represented by an output of 0V or 4mA.
Maximum %
20.00
The moisture value represented by an output of 10.00V or
20mA.
Calibrate %
0.00
Used to force a known value to the output for calibration of
external equipment.
Variable
NOW
Selects which moisture variable will be output. Options
available are...
Now
- described elsewhere.
Average
- described elsewhere.
V-Clip
- sensor voltage after rate limiting filter has been
applied.
Calibrate
- provides a fixed value determined by the Calibrate
% parameter above.
Group M1, Group M2, Group Avg
see
Grouping channels
(pon page 49).
Содержание Hydro-View HV02
Страница 20: ...20 HYDRO VIEW USER GUIDE D0022 REV 1 06 Notes ...
Страница 30: ...MATERIAL IN USE 30 HYDRO VIEW USER GUIDE D0022 REV 1 06 Notes ...
Страница 42: ...SUGGESTED WORKING VALUES 42 HYDRO VIEW USER GUIDE D0022 REV 1 06 Notes ...
Страница 48: ......
Страница 53: ...GROUPING CHANNELS HYDRO VIEW USER GUIDE D0022 REV 1 06 53 Figure 17 Four bin grouped system ...
Страница 54: ...EXAMPLE A FOUR BIN GROUPED SYSTEM USING RS485 54 HYDRO VIEW USER GUIDE D0022 REV 1 06 Notes ...
Страница 69: ...INSTALLATION HYDRO VIEW USER GUIDE D0022 REV 1 06 69 Notes ...
Страница 70: ......
Страница 80: ...ELECTROMAGNETIC COMPATIBILITY DIRECTIVE 89 336 EEC 80 HYDRO VIEW USER GUIDE D0022 REV 1 06 Notes ...
Страница 82: ...82 HYDRO VIEW USER GUIDE D0022 REV 1 06 Notes ...