SENSOR CONNECTIONS – HYDRO-PROBE I
58
HYDRO-VIEW USER GUIDE
D0022
REV 1.06
Sensor Connections – Hydro-Probe I
Hydronix moisture sensors are terminated in a 6-way Bulgin connector with a 1m lead
attached to the sensor. It is therefore necessary to wire an extension cable between the
sensor connector and the Hydro-View unit. The connection diagram is given in Figure 21.
The cable should be 5 or 6 core screened multi-core with 0.5mm
2
conductors.
Note that it is NOT satisfactory to use a 4-core cable with terminals 2 and 5 connected at
the Hydro-View end - this will lead to ‘common impedance’ type interference problems.
The +15V power rail is protected internally against short circuits by a 500mA plug-in fuse
(FS1) which may be replaced easily if required.
Warning:
Never connect other equipment to the +/-15V supply intended for the
moisture sensor.
Содержание Hydro-View HV02
Страница 20: ...20 HYDRO VIEW USER GUIDE D0022 REV 1 06 Notes ...
Страница 30: ...MATERIAL IN USE 30 HYDRO VIEW USER GUIDE D0022 REV 1 06 Notes ...
Страница 42: ...SUGGESTED WORKING VALUES 42 HYDRO VIEW USER GUIDE D0022 REV 1 06 Notes ...
Страница 48: ......
Страница 53: ...GROUPING CHANNELS HYDRO VIEW USER GUIDE D0022 REV 1 06 53 Figure 17 Four bin grouped system ...
Страница 54: ...EXAMPLE A FOUR BIN GROUPED SYSTEM USING RS485 54 HYDRO VIEW USER GUIDE D0022 REV 1 06 Notes ...
Страница 69: ...INSTALLATION HYDRO VIEW USER GUIDE D0022 REV 1 06 69 Notes ...
Страница 70: ......
Страница 80: ...ELECTROMAGNETIC COMPATIBILITY DIRECTIVE 89 336 EEC 80 HYDRO VIEW USER GUIDE D0022 REV 1 06 Notes ...
Страница 82: ...82 HYDRO VIEW USER GUIDE D0022 REV 1 06 Notes ...