HYDRO-VIEW USER GUIDE
D0022
REV 1.06
31
Advanced user options and configuration
The menu options described in this section are only available if the relevant access code
has been entered, otherwise they will appear as ... on the Hydro-View display.
Access codes are listed at the end of this manual.
Sensor calibration
Allows calibration values to be set from the results obtained by
taking laboratory samples.
Unscaled input
Displays the sensor input prior to scaling for use during
calibration.
Sensor set-up
Allows the operating parameters of the sensor to be adjusted.
Output set-up
Allows the analogue output range and variable to be adjusted.
Sensor diagnostic
Displays some engineering values about the sensor and the
Hydro-View.
Audible Alarm
Determines whether the internal bleeper is sounded when a
sensor alarm is triggered.
Material Mode
Determines how the
Material in use
number will be selected.
RS232 Usage
Configures the RS232 serial port.
RS485 Usage
Configures the (optional) RS485 port.
Data Log Rate
Sets how frequently moisture data will be sent to the serial
port(s).
Station number
Sets the station number of this Hydro-View unit in a multi-drop
configuration.
Number of slaves
In a master/slave Hydro-View configuration, determines how
many slave units are to be used.
Comms diagnostic
Displays performance information about the RS232 and RS485
communications ports.
Input 1 use
Selects how digital input 1 will be used.
Input 2 use
Selects how digital input 2 will be used.
Language
Allows the operating language of the Hydro-View to be chosen
from a list of supported languages.
Factory defaults
Allows factory default parameters to be restored.
Factory test
Executes a set of diagnostic programs on the Hydro-View to
verify operation of the unit.
Sensor calibration
The electrical properties of different materials can be quite diverse and it is often
necessary to calibrate the moisture sensors to suit different materials.
Hydro-View allows the calibration values for up to ten different materials to be stored
internally in a
material table
which can then be recalled instantly by using a
Material in use
number. This number can be selected either locally at the keypad or remotely using the
serial ports.
The Hydro-View uses a two-point calibration method which requires you to enter values
from two samples of material at different moisture contents. A minimum difference of 1%
in moisture content between the two samples is required to reduce possible calibration
errors, although a larger difference is preferable.
Parameter name
Default Value
Description
Unscaled i/p 1
4.15
Unscaled input value for calibration sample 1
Moisture % 1
0.00
Actual moisture % value for calibration sample 1
Unscaled i/p 2
28.40
Unscaled input value for calibration sample 2
Moisture % 2
11.00
Actual moisture % value for calibration sample 2
SSD %
0.00
SSD offset for this material
Содержание Hydro-View HV02
Страница 20: ...20 HYDRO VIEW USER GUIDE D0022 REV 1 06 Notes ...
Страница 30: ...MATERIAL IN USE 30 HYDRO VIEW USER GUIDE D0022 REV 1 06 Notes ...
Страница 42: ...SUGGESTED WORKING VALUES 42 HYDRO VIEW USER GUIDE D0022 REV 1 06 Notes ...
Страница 48: ......
Страница 53: ...GROUPING CHANNELS HYDRO VIEW USER GUIDE D0022 REV 1 06 53 Figure 17 Four bin grouped system ...
Страница 54: ...EXAMPLE A FOUR BIN GROUPED SYSTEM USING RS485 54 HYDRO VIEW USER GUIDE D0022 REV 1 06 Notes ...
Страница 69: ...INSTALLATION HYDRO VIEW USER GUIDE D0022 REV 1 06 69 Notes ...
Страница 70: ......
Страница 80: ...ELECTROMAGNETIC COMPATIBILITY DIRECTIVE 89 336 EEC 80 HYDRO VIEW USER GUIDE D0022 REV 1 06 Notes ...
Страница 82: ...82 HYDRO VIEW USER GUIDE D0022 REV 1 06 Notes ...