APPLICATIONS OF HYDRO-VIEW WITH HYDRONIX MOISTURE SENSORS
12
HYDRO-VIEW USER GUIDE
D0022
REV 1.06
input reading for calibration purposes which can only be done by providing the
average/hold input as described.
A
LWAYS
position the sensor so that the moisture value used is representative of the batch
being weighed. Generally, this means positioning the sensor as close to the point of
weighing as possible, within the constraints of ensuring sufficient material covers the
sensor.
A
LWAYS
perform the weight correction ‘on the fly’, towards the end of the same weighing
cycle. Some batch systems use the moisture value from one batch to correct the next
sequential batch - this is not adequate, as there may be large batch to batch changes in
moisture content. In some situations this can lead to the system performance actually
being worse when the moisture value is used.
A
LWAYS
ensure that the storage bins are managed correctly. Running out of material
during a batch may lead to errors in the reading, even though out of range readings are
not included in the average computation.
A
LWAYS
ensure that you are getting a meaningful input to the Hydro-View - there is little
point wasting time performing calibrations on the system if the sensor is covered by
stagnant material or is otherwise subject to inconsistent material flow. Verify this first
using the trend recorder display, which is a convenient way of observing the material
passing the probe.
A
LWAYS
use additional sensors where several gates are used - you should not assume
that the moisture content of the material flowing out of both gates is the same - it will not
be. The results of several Hydro-View units can be combined by use of the serial ports -
see
Grouping channels
(p49) for further details.
N
EVER
attempt to use the NOW reading to correct weights - the moisture content during a
batch can change quite dramatically. We have observed variations between 6% and 16%
during a single batch - the average value computed by the Hydro-View on the same batch
was accurate (compared with a desiccation test on a large sample of the batch) to within
0.2%.
N
EVER
attempt to use the displayed value for manual input into a batch computer for
weight correction if it is possible to interface directly. This relies on an operator to input the
value, which at best will be one batch behind and at worst will never be done.
Содержание Hydro-View HV02
Страница 20: ...20 HYDRO VIEW USER GUIDE D0022 REV 1 06 Notes ...
Страница 30: ...MATERIAL IN USE 30 HYDRO VIEW USER GUIDE D0022 REV 1 06 Notes ...
Страница 42: ...SUGGESTED WORKING VALUES 42 HYDRO VIEW USER GUIDE D0022 REV 1 06 Notes ...
Страница 48: ......
Страница 53: ...GROUPING CHANNELS HYDRO VIEW USER GUIDE D0022 REV 1 06 53 Figure 17 Four bin grouped system ...
Страница 54: ...EXAMPLE A FOUR BIN GROUPED SYSTEM USING RS485 54 HYDRO VIEW USER GUIDE D0022 REV 1 06 Notes ...
Страница 69: ...INSTALLATION HYDRO VIEW USER GUIDE D0022 REV 1 06 69 Notes ...
Страница 70: ......
Страница 80: ...ELECTROMAGNETIC COMPATIBILITY DIRECTIVE 89 336 EEC 80 HYDRO VIEW USER GUIDE D0022 REV 1 06 Notes ...
Страница 82: ...82 HYDRO VIEW USER GUIDE D0022 REV 1 06 Notes ...