Save Lock
Function Group
Menu Map
Description
Programming Codes
See Also
SYSTEM
8
This
lets you disable the save function. It prohibits the
saving of the present instrument state into a save/recall memory
register. If this function is active, an error message is displayed. An
asterisk next to the key label indicates that this function is active.
SCPI:
NONE
Analyzer:
SHSV locks the registers, SHRC unlocks the registers.
(SAVE), Security Menu
“Saving and Recalling an Instrument State” in Chapter
1
“Programming Typical Measurements” in Chapter 1
Save User Preset
Function Group
Menu Map
Description
Programming Codes
See Also
SYSTEM
8
This
lets you store the present state of operation to be used
as the PRESET state. Set the swept CW generator to the desired
operating conditions. Select Save User Preset . The display shows:
- -
User Defined Preset Saved
To activate this stored information, you must set the preset mode to
User.
SCPI:
SYSTem:PRESet:SAVE
Analyzer:
NONE
Preset Mode User
“Changing the Preset Parameters” in Chapter 1
S-2 Operating and Programming Reference
Содержание HP 8360 L Series
Страница 134: ...L 2i D u o ov P D P D cn B j g z e VJ g d m 5 ...
Страница 215: ...HP IB Menu See Also CONNECTORS HP IB Getting Started Programming in Chapter 1 H 2 Operating and Programming Reference ...
Страница 347: ...Menu Maps Menu Maps 2b 1 ...
Страница 348: ......
Страница 349: ......
Страница 350: ......
Страница 351: ......
Страница 352: ......
Страница 353: ......
Страница 354: ......
Страница 355: ......
Страница 356: ......
Страница 405: ...5 Instrument History This chapter is left blank until this manual requires changes Instrument History 5 l ...