COMMAND SUMMARY
l
l
Sets and queries the manual frequency. This controls the output
frequency in swept manual mode. The limits are START and STOP.
*RST
valueisthe same as
more
information.
l
l
Sets and queries the switch that selects either swept, CW or list
operation.
value is CW.
l
l
Sets and queries the frequency multiplier. The numeric value is
rounded to the nearest integer. This function changes mapping of
frequency parameters on input to and output from the swept CW
generator. Changing this does not affect the output frequency of
the swept CW generator. Only the displayed parameters and query
responses are changed.
The equation implied by this is :
Entered/displayed frequency (Hardware
* Multiplier ) + Offset
After
the value is 1.
l
l
Queries and turns the frequency multiplier off and on.
After
the setting is OFF.
l
l
Sets and queries the frequency offset. This function changes the
mapping of the frequency parameters on input to and output from
the swept
CW
generator. Changing this does not affect the output
frequency of the swept CW generator. Only the displayed parameters
and query responses are changed. The equation implied by this is :
Entered/displayed frequency (Hardware
* Multiplier ) + Offset
After
the value is 0.
l
l
Operating and Programming Reference
S-3 1
Содержание HP 8360 L Series
Страница 134: ...L 2i D u o ov P D P D cn B j g z e VJ g d m 5 ...
Страница 215: ...HP IB Menu See Also CONNECTORS HP IB Getting Started Programming in Chapter 1 H 2 Operating and Programming Reference ...
Страница 347: ...Menu Maps Menu Maps 2b 1 ...
Страница 348: ......
Страница 349: ......
Страница 350: ......
Страница 351: ......
Страница 352: ......
Страница 353: ......
Страница 354: ......
Страница 355: ......
Страница 356: ......
Страница 405: ...5 Instrument History This chapter is left blank until this manual requires changes Instrument History 5 l ...