E
8360 Adrs
Function Group
SYSTEM
M e n u M a p
Description
This
lets you change the HP-IB address of the swept CW
generator. Enter the address desired using the numeric entry keys
or the up/down arrow keys. The address value may be set between
0 and 30. The swept CW generator stores the address value in
non-volatile memory. The default address of the swept CW generator
is 19.
Programming Codes
SCPI:
SYSTem:COMMunicate:GPIB:ADDRess
Analyzer:
NONE
See
Also
Connectors,
HP-IB
Menu
“Instrument Addresses” in Chapter 1
“Programming Typical Measurements” in Chapter 1
Enter Corr
Function Group
POWER
M e n u M a p
Description
This
lets you enter a power correction value for a frequency
point in the flatness array. A frequency point must be entered before
a correction value can be accepted, otherwise the following error
message appears:
ERROR Must first enter correction freq.
The up/down arrow keys let you scroll through the frequency points
available for power correction. If no correction value is entered
for a frequency point, the swept CW generator display indicates
Undefined. The range of acceptable values is -40 to
An
asterisk next to the key label indicates that this feature is active.
Operating and Programming Reference E-l
Содержание HP 8360 L Series
Страница 134: ...L 2i D u o ov P D P D cn B j g z e VJ g d m 5 ...
Страница 215: ...HP IB Menu See Also CONNECTORS HP IB Getting Started Programming in Chapter 1 H 2 Operating and Programming Reference ...
Страница 347: ...Menu Maps Menu Maps 2b 1 ...
Страница 348: ......
Страница 349: ......
Страница 350: ......
Страница 351: ......
Страница 352: ......
Страница 353: ......
Страница 354: ......
Страница 355: ......
Страница 356: ......
Страница 405: ...5 Instrument History This chapter is left blank until this manual requires changes Instrument History 5 l ...