Using Detector
Detector calibration is useful for characterizing and compensating for
Calibration
negative diode detectors used in external leveling. Detectors may be
characterized by three operating regions as shown in Figure l-12:
the square law, the linear, and the transition region. The following
steps use an HP 437B to automatically characterize the operating
regions and use this information to automatically compensate for
the detector being used. The equipment setup shown in Figure 1-22
assumes that the steps necessary to correctly externally level have
been followed.
Refer to menu map 9, USER CAL.
H P - 1 6
SENSOR
Figure l-22. Automatically Characterizing and Compensating for a Detector
1. Connect the power meter as shown.
2. Zero and calibrate the power meter/sensor.
3. Enter the appropriate power sensor calibration factors into the
power meter.
4. Enable the power meter/sensor cal factor array. For operating
information on the HP 437B power meter, refer to its operating
and service manual.
5. Connect the power sensor to the output of the coupler
(or splitter).
6. On the swept CW generator, set the power level and start/stop
frequency information as desired.
7.
Press (USER).
Select Ext
Cal . The power meter is now under swept
CW generator control and is performing the sequence of steps
necessary to generate the compensation information.
Getting Started Advanced 1-47
Содержание HP 8360 L Series
Страница 134: ...L 2i D u o ov P D P D cn B j g z e VJ g d m 5 ...
Страница 215: ...HP IB Menu See Also CONNECTORS HP IB Getting Started Programming in Chapter 1 H 2 Operating and Programming Reference ...
Страница 347: ...Menu Maps Menu Maps 2b 1 ...
Страница 348: ......
Страница 349: ......
Страница 350: ......
Страница 351: ......
Страница 352: ......
Страница 353: ......
Страница 354: ......
Страница 355: ......
Страница 356: ......
Страница 405: ...5 Instrument History This chapter is left blank until this manual requires changes Instrument History 5 l ...