l
SWEep:STEP
suffix]
l
SWEep:STEP ?
I
Sets and queries the size of each frequency step.
is governed
by the equation:
STEP = SPAN/POINTS.
If you change step size then the number of points will be changed
to span/step and a
Parameter Bumped
execution error is reported.
If span or points are changed, then STEP= SPAN/POINTS. The
step sweep command creates a coupling with sweeptime also.
If the number of points is changed through this coupling and
is
ON
and
TIME:AUTOis
dwell is changed to
Span is normally an independent variable
but is changed to STEP x POINTS if both of these parameters are
changed in the same message. *RST value is
l
SWEep:TIME
suffix]
I
I
l
SWEep:TIME?
Sets and queries the current sweep time. The dwell time can be
coupled
to sweep timeifSWE:DWEL:AUTO is ON.
The
dwell time is
then governed by the equation
= SWEEPTIME/POINTS.
Changing either sweep time or the number of points causes
to be recalculated but does not cause an error. If you attempt to
change the dwell time then :AUTOis set to
OFF
then sweep time is independent of the dwell time and the number of
points. *RST value is
MIN.
l
SWEep:TIME:AUTO
l
SWEep:TIME:AUTO?
Sets and queries the automatic sweep time switch.
ON
The
value of the sweep time is automatically to
minimum.
O F F
Attempting to set a sweep time faster than allowed in the
AUTO mode causes this switch to change
to AUTO ON
even
if it was previously in the
AUTO OFF mode.
*RST state is ON.
l
SWEep:TIME:LLIMit
suffix]
l
SWEep:TIME:LLIMit?
Sets and queries the lower sweep time limit. This value specifies the
fastest sweep time that you wants the instrument to allow either on
input or when calculated internally when in
AUTO ON
mode. This
value must be greater than 10 ms. *RST value is 10 ms.
l
SWEep:TRIGger:SOURce
l
SWEep:TRIGger:SOURce?
Sets and queries the stepped sweep point-to-point trigger source.
This only applies when
SWEep:GENis set to
S-44 Operating and Programming Reference
Содержание HP 8360 L Series
Страница 134: ...L 2i D u o ov P D P D cn B j g z e VJ g d m 5 ...
Страница 215: ...HP IB Menu See Also CONNECTORS HP IB Getting Started Programming in Chapter 1 H 2 Operating and Programming Reference ...
Страница 347: ...Menu Maps Menu Maps 2b 1 ...
Страница 348: ......
Страница 349: ......
Страница 350: ......
Страница 351: ......
Страница 352: ......
Страница 353: ......
Страница 354: ......
Страница 355: ......
Страница 356: ......
Страница 405: ...5 Instrument History This chapter is left blank until this manual requires changes Instrument History 5 l ...