Step Control Master
MISTER
SLRVE
SYNTHESIZER
SYNTHESIZER
Figure S-l. Connections Required for a Two-Tone Scalar Network Analyzer Measurement System
1. Designate one swept CW generator as the master, the other as the
slave.
2. Make the connections.
3. To avoid synchronization problems, always set up the slave
(frequency and power) before setting up the master.
4. Set up the master (frequency, power, and sweep time).
5. Set the sweep time on the slave.
6. Configure the swept CW generators for step sweep, or ramp
sweep.
7. Select the appropriate triggering scheme.
8. Activate the slave mode on the slave swept CW generator.
9. Activate the master mode on the master swept CW generator.
By connecting the master’s 10 MHz reference standard to the slave’s
10 MHz reference input, the master swept CW generator’s
supplies the frequency reference for both swept CW generators.
In step sweep measurements, if the master swept CW generator
is not connected to an external controller, it must automatically
trigger the steps. Set Step
Pt Trig Auto on the master.
When a the scalar network analyzer is the step sweep controller, set
Step
Pt Trig Bus on the master swept CW generator so that
the analyzer can trigger the steps.
Programming Codes
SWEep:CONTrol:STATe
SWEep:CONTrol:TYPE
Analyzer:
NONE
See Also
Step Control Slave, Step
Menu
Operating and Programming Reference S-57
Содержание HP 8360 L Series
Страница 134: ...L 2i D u o ov P D P D cn B j g z e VJ g d m 5 ...
Страница 215: ...HP IB Menu See Also CONNECTORS HP IB Getting Started Programming in Chapter 1 H 2 Operating and Programming Reference ...
Страница 347: ...Menu Maps Menu Maps 2b 1 ...
Страница 348: ......
Страница 349: ......
Страница 350: ......
Страница 351: ......
Страница 352: ......
Страница 353: ......
Страница 354: ......
Страница 355: ......
Страница 356: ......
Страница 405: ...5 Instrument History This chapter is left blank until this manual requires changes Instrument History 5 l ...