4. While monitoring the internal detector, the RF modulator level
is varied until the detected power is equivalent to the reference
power measured in step 2.
5. Modulation is re-enabled if appropriate.
These steps are performed in approximately 200
and are repeated
any time power or frequency is changed.
See Also
listed above, Fltness Menu,
( M O D ) , [
P O W E R L E V E L
) ,
Set
“Externally Leveling the Swept CW Generator” in Chapter 1
“Working with Mixers/Reverse Power Effects” in Chapter 1
“Working with Spectrum Analyzers/Reverse Power Effects” in
Chapter 1
Operating and Programming Reference A-9