requested value. The start/stop frequencies are updated to reflect the
changes based on the coupling equations.
The swept CW generator uses “bumping” to move unspecified
frequency parameters, but if the final value of any of the frequency
headers is the result of bumping, then an error is generated since
the user is
not
getting what was specified. This means, to guarantee
sequence independence requires sending the frequency pairs in a
single message.
Example 1: (present state start 5
stop 6
20 GHZ
an error results since the
stop frequency is bumped.
22 GHZ
the final sweep does not
generate an error
(20
22).
Example 2: (present state start 5
stop = 6
22 GHZ
no error is generated,
start frequency is unchanged.
20 GHZ
still no error.
Example 3: (present state start = 5
stop = 6
20 GHZ;STOP 22 GHZ
both are fine,
22 GHZ;STARt 20 GHZ
no errors.
l
l
Sets and queries the center frequency.
The *RST value is (MAX +
l
l
?
I
l
?
Sets and queries the CW frequency. This does not change the
mode switch. *RST value is (MAX +
.
See
for more information.
l
l
l
l
Sets and queries the
center frequency coupling switch. This
switch keeps the two functions coupled together when ON. Changing
one of them, changes both. *RST setting is
OFF. See
for more information.
S-30 Operating and Programming Reference
Содержание HP 8360 L Series
Страница 134: ...L 2i D u o ov P D P D cn B j g z e VJ g d m 5 ...
Страница 215: ...HP IB Menu See Also CONNECTORS HP IB Getting Started Programming in Chapter 1 H 2 Operating and Programming Reference ...
Страница 347: ...Menu Maps Menu Maps 2b 1 ...
Страница 348: ......
Страница 349: ......
Страница 350: ......
Страница 351: ......
Страница 352: ......
Страница 353: ......
Страница 354: ......
Страница 355: ......
Страница 356: ......
Страница 405: ...5 Instrument History This chapter is left blank until this manual requires changes Instrument History 5 l ...