M
MI--M2 Sweep
Function Group
MARKER
M e n u M a p
Description
This
lets you set the swept CW generator to start sweeping
at the frequency of marker 1 (Ml), and stop sweeping at the
frequency of marker 2 (M2). M2 must have a higher frequency
value than Ml. If Ml--M2 Sweep is activated when M2 is at a
lower frequency than Ml, the values of Ml and M2 are permanently
interchanged. While this function is active, the start/stop frequencies
of the swept CW generator are changed to the values of Ml and M2.
An asterisk next to the key label indicates this feature is active.
Programming Codes
SCPI:
SWEep:MARKer:STATe
Analyzer:
function on, MPO function off.
See Also
M a r k e r
M l ,
“Marker Operation” in Chapter 1
“Programming Typical Measurements” in Chapter
Manual Sweep
Function Group
SWEEP
M e n u M a p
Description
This
lets you set the swept CW generator to the manual
sweep mode of operation. Depending on what parameter is sweeping,
you can use either the rotary knob or the ARROW keys to manually
sweep between the start/stop limits. In manual sweep mode, the
swept CW generator does not automatically retrace at the sweep
end point (the user must turn the rotary knob to retrace), and the
green SWEEP LED does not light. The resolution of the rotary
knob is 0.1% of the sweep span in either start/stop or
mode.
The resolution of the
and
arrow keys are dependent on the
Operating and Programming Reference M-l
Содержание HP 8360 L Series
Страница 134: ...L 2i D u o ov P D P D cn B j g z e VJ g d m 5 ...
Страница 215: ...HP IB Menu See Also CONNECTORS HP IB Getting Started Programming in Chapter 1 H 2 Operating and Programming Reference ...
Страница 347: ...Menu Maps Menu Maps 2b 1 ...
Страница 348: ......
Страница 349: ......
Страница 350: ......
Страница 351: ......
Страница 352: ......
Страница 353: ......
Страница 354: ......
Страница 355: ......
Страница 356: ......
Страница 405: ...5 Instrument History This chapter is left blank until this manual requires changes Instrument History 5 l ...