Enable User Flatness Correction
17. When the operation is complete, (a message is displayed) the
flatness correction array is ready to be applied to your setup.
18. Disconnect the power meter/sensor.
19. On the swept CW generator, press
[
P O W E R LEVEL
)
Where
loss
for maximum leveled power at the test
port.
20.
To save the swept CW generator parameters including the
correction table in an internal register, press (SAVE)
(n number 1 through 8).
Reactivate the HP 8757 System Interface
21. Set the analyzer to SYSINTF ON, the analyzer and swept CW
generator preset.
22. Press (RECALL)
Recall the swept CW generator parameters
from storage register 1
23. On the swept CW generator, press
(
FLTNESS O N
/
OFF
]
(amber
LED on). The power produced at the point where the power
meter/sensor was disconnected is now calibrated at the
frequencies and power level specified above.
1-46 Getting Started Advanced