Figure 1-19. Creating a User Flatness Array
For this example, refer to menu map 5, POWER.
1. The equipment setup shown in Figure 1-19 assumes that if your
setup has an external leveling configuration, the steps necessary
to correctly level have been followed. If you have questions about
external leveling, refer to “Externally Leveling the Swept CW
Generator”.
Set up Power Meter
2. Zero and calibrate the power meter/sensor.
3. Connect the power sensor to the point where flatness corrected
power is desired.
Set up Swept CW Generator Parameters
4. On the swept CW generator, press (PRESET).
5.
[
P O W E R LEVEL
]
This sets the test port power to
Create A Frequency List
6. On the swept CW generator, press FREQUENCY
Select List Menu Enter List Freq
This enters
5
as the first frequency in the list array. Entering a
frequency automatically sets the offset to 0
and the dwell to
10 ms.
8. Enter 18, 13, 11, and 20
to complete this example array.
Getting Started Advanced 1-37
Содержание HP 8360 L Series
Страница 134: ...L 2i D u o ov P D P D cn B j g z e VJ g d m 5 ...
Страница 215: ...HP IB Menu See Also CONNECTORS HP IB Getting Started Programming in Chapter 1 H 2 Operating and Programming Reference ...
Страница 347: ...Menu Maps Menu Maps 2b 1 ...
Страница 348: ......
Страница 349: ......
Страница 350: ......
Страница 351: ......
Страница 352: ......
Страница 353: ......
Страница 354: ......
Страница 355: ......
Страница 356: ......
Страница 405: ...5 Instrument History This chapter is left blank until this manual requires changes Instrument History 5 l ...