Creating a User Flatness Array, Example 2
This example shows how to use the swept CW generator and a power
meter in manual entry mode. This example also introduces two
features of the swept CW generator. The
Freq Follow
simplifies the data entry process and the
List Mode sets up
a list of arbitrary test frequencies.
The frequency follow feature automatically sets the source to a CW
test frequency equivalent to the active correction frequency in the
user flatness correction table. The front panel arrow keys are used
to move around the correction table and enter frequency-correction
pairs. Simultaneously, the swept CW generator test frequency is
updated to the selected correction frequency without exiting the
correction table.
To further simplify the data entry process, the swept CW generator
allows you to enter correction data into the user flatness correction
table by adjusting the front panel knob until the desired power
level is displayed on the power meter. The user flatness correction
algorithm automatically calculates the appropriate correction and
enters it into the table. If you already have a table of correction data
prepared, it can be entered directly into the correction table using
the front-panel keypad of the swept CW generator.
With the list mode feature, you may enter the test frequencies into
a table in any order and specify an offset (power) and/or a dwell
time for each frequency. When list mode is enabled, the swept CW
generator steps through the list of frequencies in the order entered.
The user flatness correction feature has the capability of copying and
entering the frequency list into the correction table. Since the offset
in the list mode table is not active during the user flatness correction
data entry process, the value of the correction data is determined
as if no offset is entered. When user flatness correction and list
mode (with offsets) are enabled, the swept CW generator adjusts the
output power by an amount equivalent to the sum of the correction
data and offset for each test frequency. You must make sure that the
resulting power level is still within the ALC range of the swept CW
generator.
1-36 Getting Started Advanced
Содержание HP 8360 L Series
Страница 134: ...L 2i D u o ov P D P D cn B j g z e VJ g d m 5 ...
Страница 215: ...HP IB Menu See Also CONNECTORS HP IB Getting Started Programming in Chapter 1 H 2 Operating and Programming Reference ...
Страница 347: ...Menu Maps Menu Maps 2b 1 ...
Страница 348: ......
Страница 349: ......
Страница 350: ......
Страница 351: ......
Страница 352: ......
Страница 353: ......
Страница 354: ......
Страница 355: ......
Страница 356: ......
Страница 405: ...5 Instrument History This chapter is left blank until this manual requires changes Instrument History 5 l ...