LINT2 INTERRUPT:
This can only be caused by an internal
processor error. Refer to Chapter 4 for instructions on
a
qualified service technician.
LINT6 INTERRUPT:
This can only be caused by an internal
processor error. Refer to Chapter 4 for instructions on contacting a
qualified service technician.
Number of pts must. be
2 !!:
This error occurs in association
with the user power flatness, auto fill number of points, feature. If
the number of points requested is less than two, you will see this
error message. Correct by entering number of points greater than or
equal to two.
OPTION NOT INSTALLED:
This error occurs when the HP-IB
language switch is set to a configuration requiring a certain
firmware/hardware combination to be present in the swept CW
generator. See Chapter 3 for information on language selection and
see “Specifications” for information on option available.
PRIV VIOLATION EXCEPTION:
This can only be caused by
an internal processor error. Refer to Chapter 4 for instructions on
contacting a qualified service technician.
RECALL REGISTERS LOST:
This message can appear in
association with the security menu feature, memory clear. Also, a
weak, dead, or disconnected internal battery can cause this message.
Refer to Chapter 4 for instructions on contacting a qualified service
technician.
REQUIRES system interface OFF:
This error message
indicates that the swept CW generator is connected to a network
analyzer and can not run selftest. Correct by disconnecting the
system interface cable from the swept CW generator.
SPURIOUS INTERRUPT:
This can only be caused by an internal
processor error. Refer to Chapter 4 for instructions on contacting a
qualified service technician.
SYSTEM CONTROLLER
ON
BUS:
This error message is
generated when an external controller is active on the HP-IB and
the swept CW generator has attempted to act as the controller.
Disconnect the HP-IB interface or return the swept CW generator to
LOCAL operation and repeat the request.
TOO MANY CORRECTION PTS REQUESTED:
This error
occurs in association with the user power flatness menu. The
maximum number of correction points has been reached or the
addition of the points requested will exceed the maximum. The
maximum number of points available is 801.
TOO MANY LIST POINTS REQUESTED:
This error occurs in
association with the frequency list menu. The maximum number of
list points has been reached or the addition of the points requested
will exceed the maximum. The maximum number of points available
is 801.
Error Messages
Содержание HP 8360 L Series
Страница 134: ...L 2i D u o ov P D P D cn B j g z e VJ g d m 5 ...
Страница 215: ...HP IB Menu See Also CONNECTORS HP IB Getting Started Programming in Chapter 1 H 2 Operating and Programming Reference ...
Страница 347: ...Menu Maps Menu Maps 2b 1 ...
Страница 348: ......
Страница 349: ......
Страница 350: ......
Страница 351: ......
Страница 352: ......
Страница 353: ......
Страница 354: ......
Страница 355: ......
Страница 356: ......
Страница 405: ...5 Instrument History This chapter is left blank until this manual requires changes Instrument History 5 l ...