Set up Swept CW Generator Parameters
6.
7.
8.
On the swept CW generator, press (PRESET).
FREQUENCY (START)
(STOP)
(
P O W E R L E V E L
)
Access User Flatness Correction Menu
9.
10.
Press POWER
Select
Menu.
Select Delete
Menu
Delete All . This step insures that the
flatness array is empty.
11. Press
Leave the delete menu and return to the previous
menu.
12. Enter the frequency points at which the correction information
will be taken. Choose either the point-by-point entry method
Enter Freq or the automatic frequency point generation
Auto Fill Start. For this example, select Auto Fill Start
13. Select Auto Fill Stop
Auto Fill
Notice that a frequency list starting at 4 and ending at 10
with an increment value of 1
is created.
Enter Correction Data into Array
14. Select Mtr
Menu
Measure
All . The power meter
is now under swept CW generator control and is performing
the sequence of steps necessary to generate the correction
information at each frequency point.
If an HP-IB error message is displayed, verify that the interface
connections are correct. Check the HP-IB address of the power
meter and ensure that it is the same address the swept CW
generator is using (address 13 is assumed). Refer to the menu
map 8, System, for the key sequence necessary to reach
Meter Adrs .
Enable User Flatness Correction
15.
When the operation is complete, (a message is displayed) the
flatness correction array is ready to be applied to your setup.
Disconnect the power meter/sensor and press
(
FLTNESS O N
/
OFF
)
(amber LED on). T he p ower produced at the point where the
power meter/sensor was disconnected is now calibrated at the
frequencies and power level specified above.
Getting Started Advanced 1-35
Содержание HP 8360 L Series
Страница 134: ...L 2i D u o ov P D P D cn B j g z e VJ g d m 5 ...
Страница 215: ...HP IB Menu See Also CONNECTORS HP IB Getting Started Programming in Chapter 1 H 2 Operating and Programming Reference ...
Страница 347: ...Menu Maps Menu Maps 2b 1 ...
Страница 348: ......
Страница 349: ......
Страница 350: ......
Страница 351: ......
Страница 352: ......
Страница 353: ......
Страница 354: ......
Страница 355: ......
Страница 356: ......
Страница 405: ...5 Instrument History This chapter is left blank until this manual requires changes Instrument History 5 l ...