generator is using (address 13 is assumed). Refer to the menu
map 8, System, for the key sequence necessary to reach
Meter
.
Enable User Flatness Correction
13. When the operation is complete, (a message is displayed) the
flatness correction array is ready to be applied to your setup.
14. To save the swept CW generator parameters including the
correction table in an internal register, press (SAVE)
(n = number 1 through 8).
15. Disconnect the power meter/sensor and press
(
FLTNESS O N
/
OFF
)
(amber LED on). The p ower produced at the point where the
power meter/sensor was disconnected is now calibrated at the
frequencies and power level specified above.
1-42 Getting Started Advanced