Function Group
INSTRUMENT STATE
Menu Map
NONE
Description
This
(green) causes the swept CW generator to perform a
short version of self-test, and initializes the swept CW generator to
a standard starting configuration. Two states can be defined for the
standard configuration: Factory or User.
Press
at any time to test the swept CW generator and
restore to a standard configuration. If the red LED adjacent to
the
key (labeled INSTR CHECK) stays on after preset,
the swept CW generator failed self-test; refer to
8360 B-Series
Swept Signal Generator/HP 8360 L-Series Swept CW Generator
Troubleshooting Guide.
Cycling power with the POWER switch does not have the same
effect as presetting the swept CW generator. Cycling power causes
the swept CW generator to display the programming language, the
HP-IB address, and the firmware revision date. After the swept CW
generator displays this data, it restores its configuration to the state
before power was turned off.
Programming Codes
SCPI:
SYSTem:PRESet[:EXECute]
Analyzer:
IP
See Also
Preset Mode Factory, Preset Mode User
“Changing the Preset Parameters” in Chapter 1
“Programming Typical Measurements” in Chapter 1
Operating and Programming Reference P-9
Содержание HP 8360 L Series
Страница 134: ...L 2i D u o ov P D P D cn B j g z e VJ g d m 5 ...
Страница 215: ...HP IB Menu See Also CONNECTORS HP IB Getting Started Programming in Chapter 1 H 2 Operating and Programming Reference ...
Страница 347: ...Menu Maps Menu Maps 2b 1 ...
Страница 348: ......
Страница 349: ......
Страница 350: ......
Страница 351: ......
Страница 352: ......
Страница 353: ......
Страница 354: ......
Страница 355: ......
Страница 356: ......
Страница 405: ...5 Instrument History This chapter is left blank until this manual requires changes Instrument History 5 l ...