6. Turn off the HP 8757 System Interface. Use the analyzer
SYSINTF ON OFF
found under the SYSTEM menu to
deactivate the system interface.
Access User Flatness Correction Menu
7. On the swept CW generator, press POWER
Select
Fltness Menu.
Select Delete Menu Delete All . This step insures that the
flatness array is empty.
9. Press (PRIOR). Leave the delete menu and return to the previous
soft key menu.
10. Select Auto Fill Start
Set the first frequency in
correction table to 2
11. Auto Fill Stop
Set the last frequency in
correction table to 20
12. Auto Fill
(MHz. Set the frequency increment to
every 100 MHz from 2 to 20
Set up Power Meter
13. Zero and calibrate the power meter/sensor.
14. Connect the power sensor to test port.
15. Enter and store in the power meter, the power sensor’s cal factors
for correction frequencies to be used.
Enter Correction Data into Array
16. Select Mtr
Menu
Measure
Corr All. The power meter
is now under swept CW generator control and is performing
the sequence of steps necessary to generate the correction
information at each frequency point.
If an HP-IB error message is displayed verify that the interface
connections are correct. Check the HP-IB address of the power
meter and ensure that it is the same address the swept CW
generator is using (address 13 is assumed). Refer to the menu
map 8, System, for the key sequence necessary to reach
Meter Adrs .
Getting Started Advanced l-45
Содержание HP 8360 L Series
Страница 134: ...L 2i D u o ov P D P D cn B j g z e VJ g d m 5 ...
Страница 215: ...HP IB Menu See Also CONNECTORS HP IB Getting Started Programming in Chapter 1 H 2 Operating and Programming Reference ...
Страница 347: ...Menu Maps Menu Maps 2b 1 ...
Страница 348: ......
Страница 349: ......
Страница 350: ......
Страница 351: ......
Страница 352: ......
Страница 353: ......
Страница 354: ......
Страница 355: ......
Страница 356: ......
Страница 405: ...5 Instrument History This chapter is left blank until this manual requires changes Instrument History 5 l ...