To check out the oscilloscope
1
Is there an interface module connected to the oscilloscope?
If yes, do the following steps. If not, go to step 2.
a
Turn off the oscilloscope.
b
Remove the module.
c
Turn on the oscilloscope, then check for the failing symptom.
If the failing symptom disappears, replace the module. If not, go to
step 2.
2
Disconnect any external cables from the front panel.
3
Disconnect the power cord, then remove the cover.
4
Connect the power cord, then turn on the oscilloscope.
If the display comes on after a few seconds, (HP logo and copyright text,
followed by a graticule with text at top of the display) go to "To check the
Low Voltage Power Supply," on page 4–45. If after checking the Low Voltage
Power Supply the voltages are within the test limits, go to step 8. If not, go to
step 6. If the display did not come on, do the steps below.
a
Check the intensity knob setting to see if it
'
s set too low.
b
If there is still no display, disconnect the power cord.
c
Check all cable connections.
d
Go to "To check the Low Voltage Power Supply," on page 4–45.
If the voltages are within the limits go to step 5. If not, go to step 6.
Service
Troubleshooting the Oscilloscope
4–42
Содержание HP 54600B
Страница 7: ...vi ...
Страница 12: ...1 The Oscilloscope at a Glance ...
Страница 16: ...Softkey Menu Reference Press this key To obtain this menu Figure 1 4 To obtain this menu Press this key 1 5 ...
Страница 28: ...2 Operating Your Oscilloscope ...
Страница 69: ...2 42 ...
Страница 70: ...3 Using Option 005 Enhanced TV Video Trigger HP 54602B ...
Страница 148: ...Exploded view of oscilloscope Figure 4 11 Service Replacing Parts in the Oscilloscope 4 59 ...
Страница 151: ...Exploded view of Option 005 and related oscilloscope parts Figure 4 12 Service Replacing Parts in the Oscilloscope 4 62 ...
Страница 153: ...4 64 ...
Страница 154: ...5 Performance Characteristics ...
Страница 175: ...Glossary 8 ...