To verify horizontal
∆
t and 1/
∆
t accuracy
In this test you verify the horizontal
∆
t and 1/
∆
t accuracy by measuring the
output of a time mark generator with the oscilloscope.
Test limits
±
0.01%
±
0.2% of full scale
±
200 ps.
Table 4-9
Equipment Required
Equipment
Critical specifications
Recommended Model/Part
Time marker generator
Stability 5 ppm after 1/2 hour
TG 501A and TM 503A
Cable
BNC
HP 10503A
Termination
50
Ω
, BNC connectors (m) (f)
HP 10100C
1
Connect the time mark generator to channel 1 using the feedthrough
at the oscilloscope input. Then, set the time mark generator for 0.1
ms markers.
2
Setup the oscilloscope.
a
Press
Setup
, then press the
Default Setup
softkey.
b
Press
Diplay
, then press the
Vectors Off
softkey.
c
Press
Autoscale
.
d
Set the time base to 20
µ
s/div.
e
Adjust the trigger level to obtain a stable display.
Service
Verifying Oscilloscope Performance
4–16
Содержание HP 54600B
Страница 7: ...vi ...
Страница 12: ...1 The Oscilloscope at a Glance ...
Страница 16: ...Softkey Menu Reference Press this key To obtain this menu Figure 1 4 To obtain this menu Press this key 1 5 ...
Страница 28: ...2 Operating Your Oscilloscope ...
Страница 69: ...2 42 ...
Страница 70: ...3 Using Option 005 Enhanced TV Video Trigger HP 54602B ...
Страница 148: ...Exploded view of oscilloscope Figure 4 11 Service Replacing Parts in the Oscilloscope 4 59 ...
Страница 151: ...Exploded view of Option 005 and related oscilloscope parts Figure 4 12 Service Replacing Parts in the Oscilloscope 4 62 ...
Страница 153: ...4 64 ...
Страница 154: ...5 Performance Characteristics ...
Страница 175: ...Glossary 8 ...