5
Disconnect the display cable, then check the following signals on the
system board.
Table 4-18
Signals at U56
Signal
Frequency
Pulse width
Voltage
U16 Pin 7
DE
19.72 kHz
38.0
µ
s
2.6 Vp-p
U16 Pin 24
Hsync
19.72 kHz
3.0
µ
s
5.0 Vp-p
U9 Pin 2
Vsync
60.00 Hz
253.5
µ
s
5.2 Vp-p
If the signals are good, replace the display assembly. If not, replace the
system board.
6
Disconnect the Low Voltage Power Supply ribbon cable from the
display board.
7
Measure the power supply voltages again (steps 1-3).
If the voltages are within the test limits, replace the display assembly. If not,
do the steps below.
a
Disconnect the power cord.
b
Disconnect the ribbon cable from the power supply.
c
Connect the dummy load to the power supply connector.
d
Connect the power cord, then measure the power supply voltages
again (see new tolerances below).
+
5.1 V (4.95 V to
+
5.25 V)
+
15.75 V (
+
15 V to
+
16.5 V)
−
15.75 V (
−
15 V to
−
16.5 V)
If the voltages are now within the test limits, replace the system
board. If not, replace the power supply.
Service
Troubleshooting the Oscilloscope
4–43
Содержание HP 54600B
Страница 7: ...vi ...
Страница 12: ...1 The Oscilloscope at a Glance ...
Страница 16: ...Softkey Menu Reference Press this key To obtain this menu Figure 1 4 To obtain this menu Press this key 1 5 ...
Страница 28: ...2 Operating Your Oscilloscope ...
Страница 69: ...2 42 ...
Страница 70: ...3 Using Option 005 Enhanced TV Video Trigger HP 54602B ...
Страница 148: ...Exploded view of oscilloscope Figure 4 11 Service Replacing Parts in the Oscilloscope 4 59 ...
Страница 151: ...Exploded view of Option 005 and related oscilloscope parts Figure 4 12 Service Replacing Parts in the Oscilloscope 4 62 ...
Страница 153: ...4 64 ...
Страница 154: ...5 Performance Characteristics ...
Страница 175: ...Glossary 8 ...