To capture a single event
To capture a single event, you need some previous knowledge of the signal in
order to set up the trigger level and slope. For example, if the event is
derived from TTL logic, a trigger level of 2 volts should work on a rising edge.
The following steps show you how to use the oscilloscope to capture a single
event.
1
Connect a signal to the oscilloscope.
2
Set up the trigger.
•
Press
Source
. Select a trigger source with the softkeys.
•
Press
Slope/Coupling
. Select a trigger slope with the softkeys.
•
Turn the Level knob to a point where you think the trigger should work.
3
Press
Mode
, then press the
Single
softkey.
4
Press
Erase
to clear previous measurements from the display.
5
Press
Run
.
Pressing
Run
arms the trigger circuit. When the trigger conditions
are met, data appears on the display representing the data points that
the oscilloscope obtained with one acquisition. Pressing the Run key
again rearms the trigger circuit and erases the display.
Operating Your Oscilloscope
To capture a single event
2–8
Содержание HP 54600B
Страница 7: ...vi ...
Страница 12: ...1 The Oscilloscope at a Glance ...
Страница 16: ...Softkey Menu Reference Press this key To obtain this menu Figure 1 4 To obtain this menu Press this key 1 5 ...
Страница 28: ...2 Operating Your Oscilloscope ...
Страница 69: ...2 42 ...
Страница 70: ...3 Using Option 005 Enhanced TV Video Trigger HP 54602B ...
Страница 148: ...Exploded view of oscilloscope Figure 4 11 Service Replacing Parts in the Oscilloscope 4 59 ...
Страница 151: ...Exploded view of Option 005 and related oscilloscope parts Figure 4 12 Service Replacing Parts in the Oscilloscope 4 62 ...
Страница 153: ...4 64 ...
Страница 154: ...5 Performance Characteristics ...
Страница 175: ...Glossary 8 ...