4
Press
Voltage
, then press the
Vp-p
softkey.
Wait a few seconds for the measurement to settle (averaging is complete),
then note the Vp-p reading from the bottom of the display.
Vp-p = _______ mV.
5
Set the calibration factor percent of the power meter to the 1 MHz
value from the calibration chart on the probe, then press dB (REF)
on the power meter to set a 0 dB reference.
6
Change the frequency of the signal generator to the value shown
below for your instrument.
Table 4-8
Signal Generator Frequency Setting
Selected
Channel
HP 54600B
HP 54601B
HP 54602B *
HP 54603B
Channel 1
100 MHz
100 MHz
150 MHz
60 MHz
Channel 2
100 MHz
100 MHz
150 MHz
60 MHz
Channel 3
—
100
MHz
250
MHz
—
Channel 4
—
100
MHz
250
MHz
—
*1mv/div to 5 mv/div = 100 MHz.
7
Set the calibration factor of the power meter to 100 MHz percent
value from the chart on the probe.
Adjust the amplitude of the signal generator for a power reading as close as
possible to 0.0 dB (REL). Power meter reading = ______ dB.
Service
Verifying Oscilloscope Performance
4–14
Содержание HP 54600B
Страница 7: ...vi ...
Страница 12: ...1 The Oscilloscope at a Glance ...
Страница 16: ...Softkey Menu Reference Press this key To obtain this menu Figure 1 4 To obtain this menu Press this key 1 5 ...
Страница 28: ...2 Operating Your Oscilloscope ...
Страница 69: ...2 42 ...
Страница 70: ...3 Using Option 005 Enhanced TV Video Trigger HP 54602B ...
Страница 148: ...Exploded view of oscilloscope Figure 4 11 Service Replacing Parts in the Oscilloscope 4 59 ...
Страница 151: ...Exploded view of Option 005 and related oscilloscope parts Figure 4 12 Service Replacing Parts in the Oscilloscope 4 62 ...
Страница 153: ...4 64 ...
Страница 154: ...5 Performance Characteristics ...
Страница 175: ...Glossary 8 ...