To run the internal self-tests
1
Perform the keyboard test.
a
Press
Print/Utility
.
b
Press the
Self Tst
softkey, then press the
Keyboard
softkey.
A pictorial diagram of the front panel will appear on the display.
c
Press each key, and notice that when you press a key a corresponding
block on the display fills in.
d
Rotate the knobs (except the intensity) and notice that an arrow
appears on the display that points in the direction you rotate the knob.
e
Do all the keys and knobs work?
If yes, Press the
Stop
softkey two or three times (the display indicates
how many times), then go to step 2. If not, replace the keyboard and
keyboard assembly.
2
Check the output level of the
DAC
.
a
Press the
DAC
softkey.
b
Connect a multimeter to the rear panel
DC CALIBRATOR
connector.
The multimeter should read 0 V
±
500
µ
V.
c
Press any key to continue.
The multimeter should read 5 V
±
10 mV.
d
Are the DAC voltages correct?
If yes, press any key to continue. If not, replace the system board.
3
Perform the
ROM
test
a
Press the
ROM
softkey.
b
Does the display message say
Test Passed
?
If yes, press any key to continue. If not, (the display message says
Test Failed
) replace the system board.
Service
Troubleshooting the Oscilloscope
4–46
Содержание HP 54600B
Страница 7: ...vi ...
Страница 12: ...1 The Oscilloscope at a Glance ...
Страница 16: ...Softkey Menu Reference Press this key To obtain this menu Figure 1 4 To obtain this menu Press this key 1 5 ...
Страница 28: ...2 Operating Your Oscilloscope ...
Страница 69: ...2 42 ...
Страница 70: ...3 Using Option 005 Enhanced TV Video Trigger HP 54602B ...
Страница 148: ...Exploded view of oscilloscope Figure 4 11 Service Replacing Parts in the Oscilloscope 4 59 ...
Страница 151: ...Exploded view of Option 005 and related oscilloscope parts Figure 4 12 Service Replacing Parts in the Oscilloscope 4 62 ...
Страница 153: ...4 64 ...
Страница 154: ...5 Performance Characteristics ...
Страница 175: ...Glossary 8 ...