8
Change the time base to 5 ns/div.
Wait a few seconds for the measurement to settle (averaging is complete),
then note the Vp-p reading from the bottom of the display.
Vp-p = ______ mV.
9
Calculate the response using the following formula.
20 log
10
step
8
result
step
4
result
10
Correct the result from step 9 with any difference in the power meter
reading from step 7. Make sure you observe all number signs.
For example
Result from step 9 =
−
2.3 dB
Power meter reading from step 7 =
−
0.2 dB (REL)
True response = (
−
2.3)
−
(
−
0.2) =
−
2.1 dB
The true response should be
≤±
3 dB.
If the result is not
≤±
3 dB, see "Troubleshooting the Oscilloscope," on
page 4–40.
11
Repeat steps 1 to 10 for channel 2 (channels 2 to 4 on the HP 54601B
and HP 54602B).
When you measure the bandwidth on channels 3 and 4, use the 0.1 V/Div
range.
Service
Verifying Oscilloscope Performance
4–15
Содержание HP 54600B
Страница 7: ...vi ...
Страница 12: ...1 The Oscilloscope at a Glance ...
Страница 16: ...Softkey Menu Reference Press this key To obtain this menu Figure 1 4 To obtain this menu Press this key 1 5 ...
Страница 28: ...2 Operating Your Oscilloscope ...
Страница 69: ...2 42 ...
Страница 70: ...3 Using Option 005 Enhanced TV Video Trigger HP 54602B ...
Страница 148: ...Exploded view of oscilloscope Figure 4 11 Service Replacing Parts in the Oscilloscope 4 59 ...
Страница 151: ...Exploded view of Option 005 and related oscilloscope parts Figure 4 12 Service Replacing Parts in the Oscilloscope 4 62 ...
Страница 153: ...4 64 ...
Страница 154: ...5 Performance Characteristics ...
Страница 175: ...Glossary 8 ...