To analyze video waveforms
Enhanced TV/Video Trigger (HP 54602B only)
This section discusses basic TV video triggering. If you have Option 005
Enhanced TV/Video Trigger installed in your HP 54602B oscilloscope, refer to
Chapter 3 "Using Option 005 Enhanced TV/Video Trigger."
The TV sync separator in the oscilloscope has an internal clamp circuit. This
removes the need for external clamping when you are viewing unclamped
video signals. TV triggering requires two vertical divisions of display, either
channel 1 or channel 2 as the trigger source, and the selection of internal
trigger. Turning the trigger level knob in TV trigger does not change the
trigger level because the trigger level is automatically set to the sync pulse
tips.
For this exercise we connected the oscilloscope to the video output terminals
on a television. Then we set up the oscilloscope to view these parts of a TV
signal: the second vertical interval with delayed sweep windowed on the
vertical interval test signals (VITS) and the IRE displayed full screen.
1
Connect a TV signal to channel 1, then press
Autoscale
.
2
Press
Display
, then press the
Peak Det
softkey.
3
Press
Mode
, then press the
TV
softkey.
4
Press
Slope/Coupling
, then press the
Field 2
softkey.
Polarity Selects either positive or negative sync pulses.
Field 1
Triggers on the field 1 portion of the video signal.
Field 2
Triggers on the field 2 portion of the video signal.
Line
Triggers on all the TV line sync pulses.
HF Rej
Controls a 500 kHz low pass filter in the trigger path.
Operating Your Oscilloscope
To analyze video waveforms
2–38
Содержание HP 54600B
Страница 7: ...vi ...
Страница 12: ...1 The Oscilloscope at a Glance ...
Страница 16: ...Softkey Menu Reference Press this key To obtain this menu Figure 1 4 To obtain this menu Press this key 1 5 ...
Страница 28: ...2 Operating Your Oscilloscope ...
Страница 69: ...2 42 ...
Страница 70: ...3 Using Option 005 Enhanced TV Video Trigger HP 54602B ...
Страница 148: ...Exploded view of oscilloscope Figure 4 11 Service Replacing Parts in the Oscilloscope 4 59 ...
Страница 151: ...Exploded view of Option 005 and related oscilloscope parts Figure 4 12 Service Replacing Parts in the Oscilloscope 4 62 ...
Страница 153: ...4 64 ...
Страница 154: ...5 Performance Characteristics ...
Страница 175: ...Glossary 8 ...