To verify bandwidth (alternate method)
In this test you verify the bandwidth of the oscilloscope by using a power
meter and power sensor to set the output of a signal generator at 1 MHz and
the upper bandwidth limit. You use the peak-to-peak voltage at 1 MHz and
the upper bandwidth limit to calculate the bandwidth response of the
oscilloscope.
Test limits
HP 54600B and HP 54601B, all channels (
−
3 dB)
dc to 100 MHz
ac coupled 10 Hz to 100 MHz.
HP 54602B
Channels 1 & 2 (–3 dB) *
dc to 150 MHz
ac coupled 10 Hz to 150 MHz.
Channels 3 & 4 (–3 dB)
dc to 250 MHz.
HP 54603B, all channels (
−
3 dB)
dc to 60 MHz
ac coupled 10 Hz to 60 MHz.
Table 4-7
Equipment Required
Equipment
Critical specifications
Recommended Model/Part
Signal generator
1 to 100 MHz at 200 mV
(250 MHz for HP 54602B)
HP 8656B opt 001
Power meter and
Power Sensor
1 to 100 MHz
±
3% accuracy
(250 MHz for HP 54602B)
HP 436A and HP 8482A
Power splitter
Outputs differ by <0.15 dB
HP 11667B
Cable
Type N (m), 24 inch
HP 11500B
Adapter
Type N (m) to BNC (m)
HP 1251-0082
Feedthrough
50
Ω
, BNC (m) and (f)
HP 10100C
*
1mv/div to 5 mv/div dc = 100 MHz.
Service
Verifying Oscilloscope Performance
4–12
Содержание HP 54600B
Страница 7: ...vi ...
Страница 12: ...1 The Oscilloscope at a Glance ...
Страница 16: ...Softkey Menu Reference Press this key To obtain this menu Figure 1 4 To obtain this menu Press this key 1 5 ...
Страница 28: ...2 Operating Your Oscilloscope ...
Страница 69: ...2 42 ...
Страница 70: ...3 Using Option 005 Enhanced TV Video Trigger HP 54602B ...
Страница 148: ...Exploded view of oscilloscope Figure 4 11 Service Replacing Parts in the Oscilloscope 4 59 ...
Страница 151: ...Exploded view of Option 005 and related oscilloscope parts Figure 4 12 Service Replacing Parts in the Oscilloscope 4 62 ...
Страница 153: ...4 64 ...
Страница 154: ...5 Performance Characteristics ...
Страница 175: ...Glossary 8 ...