To display a signal automatically
The oscilloscope has an Autoscale feature that automatically sets up the
oscilloscope to best display the input signal. Using Autoscale requires signals
with a frequency greater than or equal to 50 Hz and a duty cycle greater than
1%.
When you press the Autoscale key, the oscilloscope turns on and scales all
channels that have signals applied, and it selects a time base range based on
the trigger source. The trigger source selected is the highest numbered input
that has a signal applied. (If a signal is connected to the external trigger
input on the HP 54600B and HP 54603B, then it is selected as the trigger
source.)
1
Connect a signal to the oscilloscope.
2
Press
Autoscale
.
When you press the Autoscale key, the oscilloscope changes the front-panel
setup to display the signal. However, if you pressed the Autoscale key
unintentionally, you can use the Undo Autoscale feature. To use this feature,
perform the following step.
•
Press
Setup
. Next, press the
Undo Autoscale
softkey.
The oscilloscope returns to the configuration in effect before you
pressed the Autoscale key.
The Oscilloscope at a Glance
To display a signal automatically
1–9
Содержание HP 54600B
Страница 7: ...vi ...
Страница 12: ...1 The Oscilloscope at a Glance ...
Страница 16: ...Softkey Menu Reference Press this key To obtain this menu Figure 1 4 To obtain this menu Press this key 1 5 ...
Страница 28: ...2 Operating Your Oscilloscope ...
Страница 69: ...2 42 ...
Страница 70: ...3 Using Option 005 Enhanced TV Video Trigger HP 54602B ...
Страница 148: ...Exploded view of oscilloscope Figure 4 11 Service Replacing Parts in the Oscilloscope 4 59 ...
Страница 151: ...Exploded view of Option 005 and related oscilloscope parts Figure 4 12 Service Replacing Parts in the Oscilloscope 4 62 ...
Страница 153: ...4 64 ...
Страница 154: ...5 Performance Characteristics ...
Страница 175: ...Glossary 8 ...